Avaliação da anestesia com baixo fluxo de gases, comparando ventilação com pressão controlada à ventilação com tempo controlado - pressão limitada e fluxo constante de gases. Modelo experimental em coelhos



Título del documento: Avaliação da anestesia com baixo fluxo de gases, comparando ventilação com pressão controlada à ventilação com tempo controlado - pressão limitada e fluxo constante de gases. Modelo experimental em coelhos
Revue: Revista brasileira de anestesiologia
Base de datos: PERIÓDICA
Número de sistema: 000302676
ISSN: 0034-7094
Autores: 1
1
Instituciones: 1Universidade de Sao Paulo, Faculdade de Medicina, Sao Paulo. Brasil
Año:
Periodo: Ene-Feb
Volumen: 53
Número: 1
Paginación: 25-38
País: Brasil
Idioma: Portugués
Tipo de documento: Artículo
Enfoque: Experimental
Resumen en español JUSTIFICATIVA Y OBJETIVOS: No obstante sean ampliamente conocidas las ventajas que envuelven sistemas de anestesia con reinhalación y bajo flujo, estos aun son poco empleados en animales de pequeño porte y en pediatría, principalmente por la carencia de equipamientos seguros para oferta de pequeño volumen corriente, por la necesidad de mayor monitoración y hasta por inseguridad de los anestesistas. El modelo experimental de este estudio fue delineado, buscándose evaluar la viabilidad del sistema con reinhalación y flujo de gases entre 500 y 1.000 ml.min-1 en animales de pequeño porte, bien como comparar la ventilación en los modos tiempo controlado - presión limitada y flujo constante de gases, con la presión controlada (PCV), en un nuevo equipamiento desarrollado para la anestesia. MÉTODO: Dieciséis conejos (hembras) fueron distribuidos aleatoriamente en dos grupos (GI y GII), anestesiados con halotano, y sometidos a ventilación controlada mecánica. En GI, se permitían ajustes en los parámetros ventilatorios del ventilador (presión de pico inspiratorio, frecuencia respiratoria y tiempo inspiratorio), de forma a mantener normocapnia, volumen corriente entre 6 y 8 ml.kg-1, y pH sanguíneo arterial entre 7,35 y 7,45. En GII, los ajustes eran realizados solamente en el momento control. RESULTADOS: Severificó reinhalación de dióxido de carbono en ambos grupos al evaluarse el sistema con reinhalación, independiente del modo de ventilación. El pH sanguíneo arterial se mantuvo dentro de los parámetros fisiológicos en el grupo GI, y en el GII se verificó acidosis respiratoria al evaluarse el sistema con reinhalación durante la ventilación con presión controlada. CONCLUSIONES: Con los resultados obtenidos se concluye que este sistema circular, en flujos entre 500 y 1.000 ml.min-1, es una alternativa en anestesia en pacientes de bajo peso, desde que la monitorización
Resumen en inglés BACKGROUND AND OBJECTIVES: Despite the well-known advantages of rebreathing systems and low flow anesthesia, they are seldom used in small animals and pediatric anesthesia. The experimental model of this study was designed to evaluate a rebreathing system with flows between 500 to 1,000 ml.min-1 in small animals, in addition to comparing time-controlled pressure-limited and constant gas flow ventilation to pressure-controlled ventilation in a new anesthesia machine. METHODS: Sixteen female rabbits were randomly allocated in two groups (GI and GII), anesthetized with halothane and submitted to mechanically-controlled ventilation. Adjustments in ventilatory parameters were allowed in GI (peak inspiratory pressure, respiratory rate and inspiratory time) to maintain normocapnia, tidal volume between 6 and 8 ml.kg-1 and arterial pH between 7.35 and 7.45. Adjustments in GII were only allowed at control moment. RESULTS: Carbon dioxide rebreathing was observed in both groups when rebreathing systems were evaluated, regardless of the ventilation mode. Arterial pH was maintained within physiologic parameters in GI, and respiratory acidosis was observed in GII when the rebreathing system was evaluated during pressure-controlled ventilation. CONCLUSIONS: According to our results, a circle system with flows between 500 and 1,000 ml.min-1 may be an alternative for low weight patients anesthesia, provided necessary monitoring is performed. Ventilation efficiency was primarily a function of adjusted ventilatory parameters, of airways compliance and resistance and not so much of the ventilation mode
Resumen en portugués JUSTIFICATIVA E OBJETIVOS: Embora sejam amplamente conhecidas as vantagens que envolvem sistemas de anestesia com reinalação e baixo fluxo, estes ainda são pouco empregados em animais de pequeno porte e em pediatria. O modelo experimental deste estudo foi delineado, procurando-se avaliar a viabilidade do sistema com reinalação e fluxo de gases entre 500 e 1.000 ml.min-1 em animais de pequeno porte, bem como comparar a ventilação nos modos tempo controlado - pressão limitada e fluxo constante de gases, com a pressão controlada (PCV), em um novo equipamento desenvolvido para anestesia. MÉTODO: Dezesseis coelhos (fêmeas) foram distribuídos aleatoriamente em dois grupos (GI e GII), anestesiados com halotano, e submetidos à ventilação controlada mecânica. Em GI, permitiam-se ajustes nos parâmetros ventilatórios do ventilador (pressão de pico inspiratório, freqüência respiratória e tempo inspiratório), de forma a manter normocapnia, volume corrente entre 6 e 8 ml.kg-1, e pH sangüíneo arterial entre 7,35 e 7,45. Em GII, os ajustes eram realizados somente no momento controle. RESULTADOS: Verificou-se reinalação de dióxido de carbono em ambos os grupos ao se avaliar o sistema com reinalação, independente do modo de ventilação. O pH sangüíneo arterial manteve-se dentro dos parâmetros fisiológicos no grupo GI, e no GII verificou-se acidose respiratória ao se avaliar o sistema com reinalação durante a ventilação com pressão controlada. CONCLUSÕES: Com os resultados obtidos conclui-se que este sistema circular, em fluxos entre 500 e 1.000 ml.mi
Disciplinas: Medicina
Palabras clave: Medicina experimental,
Conejos,
Control de calidad,
Equipo médico,
Anestesia
Keyword: Medicine,
Experimental medicine,
Rabbits,
Quality control,
Medical equipment,
Anesthesia
Texte intégral: Texto completo (Ver HTML)