Spanish abstract | Objetivo: Evaluar la precisión de Raypex 6 para localizar el foramen y ubicarse en la zona cemento dentina conducto (CDC) en conductos de molares inferiores por medio de diafanización. Material y métodos: 52 conductos permeables de 20 molares inferiores extraídos inmersos en alginato fueron utilizados. Se realizó abertura coronaria, localización, permeabilización e irrigación con hipoclorito de sodio al 5.25%. Con el localizador electrónico Raypex 6 se obtuvo conductometría electrónica. Se introdujo lima tipo K #15 o 20 en cada muestra que tenía el clip labial inserto en alginato. La pantalla del dispositivo indicó la posición del foramen apical en la barra roja y se procedió a reajustar la posición de la lima K en las dos primeras barras amarillas y la lima se fijó con resina acrílica. Los dientes se diafanizaron por medio de la técnica de ácido nítrico y se mantuvieron en salicilato de metilo. Las muestras se analizaron con microscopio clínico a 16x y de manera subjetiva se asignó el valor de preciso, si la punta de la lima se ubicó entre 0 a -0.5 mm, fuera o positivo (+) si la lima estuvo +0.1 mm o más y corto o negativo (-) si fue de -0.51 mm o menos con respecto al foramen apical. Resultados: De las 52 muestras analizadas, se encontraron 40 precisas, siete largas y cinco cortas. La estadística descriptiva demostró 76.9% de precisión. Conclusión: La longitud de trabajo electrónica con Raypex 6 mostró una adecuada precisión en conductos mesiales de molares inferiores |
English abstract | Objective: To assess precision of Raypex 6 to locate foramen and establish placement in the canal -dentin-cement (CDC) area in lower molar canals by means of diaphanization. Material and methods: 52 permeable canals from 20 extracted lower molars immersed in alginate were used. Crown opening, location, permeabilization and irrigation with 5.25% sodium hypochlorite were performed. Electronic conductometry was obtained with Raypex 6 electronic locator. A K #15/20 file type was inserted in all samples which had labial clip inserted in alginate. The device’s screen indicated position of the apical foramen in the red bar, K files position was readjusted in the first two yellow bars, the file was then fixated with acrylic resin. Teeth were diaphanized by means of the nitric acid technique; teeth were kept in methyl salicylate. Samples were analyzed with clinical microscope at 16x, they were subjectively assigned a value called precise when the tip of the file was located at 0-0.5 mm, external or positive (+), when the file was +0.1 mm or more, and short or negative (-) when it was -0.51 mm or less with respect to the apical foramen. Results: Of the 52 analyzed samples, 40 were found to be precise, seven long and five short. Descriptive statistics showed 76.9% precision. Conclusion: Electronic working length with Raypex 6 showed suitable precision in mesial canals of lower molars |