Journal: | RB. Radiologia brasileira |
Database: | PERIÓDICA |
System number: | 000301831 |
ISSN: | 0100-3984 |
Authors: | Ros, Renato Assenci1 Caldas, Linda V.E |
Institutions: | 1Comissao Nacional de Energia Nuclear, Instituto de Pesquisas Energeticas e Nucleares, Piracicaba, Sao Paulo. Brasil |
Year: | 2001 |
Season: | Mar-Abr |
Volumen: | 34 |
Number: | 2 |
Pages: | 89-94 |
Country: | Brasil |
Language: | Portugués |
Document type: | Artículo |
Approach: | Descriptivo |
English abstract | We present the quality control tests performed on a monophasic X-ray equipment and on two noninvasive monitoring systems. The results were compared with the values obtained using an invasive monitoring system and a hyperpure germanium detector, providing correction factors for utilization of the noninvasive voltage meters in the quality control program of X-ray equipment. We concluded that there is a need for periodic calibration and control |
Portuguese abstract | Este trabalho apresenta os testes de controle de qualidade feitos em um equipamento de raios X monofásico e em dois sistemas de monitoração não-invasivos. Os resultados foram comparados com os valores apresentados por um sistema de monitoração invasivo e por um detector de germânio hiperpuro, fornecendo fatores de correção para utilização dos medidores de tensão não-invasivos no programa de controle de qualidade de equipamentos de raios X, mostrando a necessidade de calibração e controle periódico |
Disciplines: | Medicina |
Keyword: | Diagnóstico, Radiaciones, Control de calidad, Rayos X, Cámaras de ionización |
Keyword: | Medicine, Diagnosis, Quality control, X-rays, Ionization chambers |
Full text: | Texto completo (Ver HTML) |