Journal: | Quimica nova |
Database: | PERIÓDICA |
System number: | 000196309 |
ISSN: | 0100-4042 |
Authors: | Queenie Siu Hang Chui1 Zucchini, Ricardo R Lichtig, Jaim |
Institutions: | 1Universidade de Sao Paulo, Instituto de Quimica, Sao Paulo. Brasil 2Instituto de Pesquisas Tecnologicas do Estado de Sao Paulo, Divisao de Quimica, Sao Paulo. Brasil |
Year: | 2001 |
Season: | May-Jun |
Volumen: | 24 |
Number: | 3 |
Pages: | 374-380 |
Country: | Brasil |
Language: | Portugués |
Document type: | Artículo |
Approach: | Experimental |
English abstract | Calculation of uncertainty of results represents the new paradigm in the area of the quality of measurements in laboratories. The guidance on the Expression of Uncertainty in Measurement of the ISO / International Organization for Standardization assumes that the analyst is being asked to give a parameter that characterizes the range of the values that could reasonably be associated with the result of the measurement. In practice, the uncertainty of the analytical result may arise from many possible sources: sampling, sample preparation, matrix effects, equipments, standards and reference materials, among others. This paper suggests a procedure for calculation of uncertainties components of an analytical result due to sample preparation (uncertainty of weights and volumetric equipment) and instrument analytical signal (calibration uncertainty). A numerical example is carefully explained based on measurements obtained for cadmium determination by flame atomic absorption spectrophotometry. Results obtained for components of total uncertainty showed that the main contribution to the analytical result was the calibration procedure |
Disciplines: | Química |
Keyword: | Química analítica, Cadmio, Espectrofotometría, Espectrofotometría de absorción atómica, Incertidumbre, Calibración, Estandarización |
Keyword: | Chemistry, Analytical chemistry, Spectrophotometry, Atomic absorption spectrophotometry, Flame atomic absorption spectrophotometry, Uncertainty, Standarization, Cadmium, Calibration |
Full text: | Texto completo (Ver HTML) |