Quadrature Slotted Surface Coil Pair for Magnetic Resonance Imaging at 4 Tesla: Phantom Study



Document title: Quadrature Slotted Surface Coil Pair for Magnetic Resonance Imaging at 4 Tesla: Phantom Study
Journal: Ingeniería. Investigación y tecnología
Database: PERIÓDICA
System number: 000348201
ISSN: 1405-7743
Authors: 1
2
1
1
Institutions: 1Universidad Autónoma Metropolitana, Departamento de Ingeniería Eléctrica, Iztapalapa, Distrito Federal. México
2Brookhaven National Laboratory, Medical Department, Upton, Nueva York. Estados Unidos de América
Year:
Season: Ene-Mar
Volumen: 13
Number: 1
Pages: 69-74
Country: México
Language: Inglés
Document type: Artículo
Approach: Aplicado, descriptivo
Spanish abstract Se presenta un arreglo de antenas transceptoras operado en cuadratura, se construyó con dos antenas superficiales con 6 ranuras circulares y separadas por un ángulo de 90°. El campo magnético de este arreglo de antenas se simuló de manera numérica a la frecuencia de 170 MHz, empleando el método de elemento finito. El arreglo de antenas se desarrolló para aplicaciones de la imagenología del cerebro por resonancia magnética a la frecuencia de 170 MHz. Las simulaciones numéricas mostraron que la interacción electromagnética entre los elementos es nula, y muestra una buena uniformidad de campo. Se generaron imágenes de fantoma con el arreglo de antenas, usando secuencias de pulsos estándar con un sistema de 4 Tesla dedicado para hacer investigación. Las imágenes en vitro muestran la factibilidad de este tipo de arreglos de antenas con secuencias de pulsos estándar para imagenología por resonancia magnética de altos campos
English abstract A coil array was composed of two slotted surface coils forming a structure with two plates at 90°, each one having 6 circular slots and is introduced in this paper. Numerical simulations of the magnetic field of this coil array were performed at 170 MHz using the finite element method to study its behaviour. This coil array was developed for brain magnetic resonance imaging to be operated at the resonant frequency of 170 MHz in the transceiver mode and quadrature driven. Numerical simulations demonstrated that electromagnetic interaction between the coil elements is negligible, and that the magnetic field showed a good uniformity. Phantom images were acquired with our coil array and standard pulse sequences on a research–dedicated 4 Tesla scanner. In vitro images showed the feasibility of this coil array for standard pulses and high field magnetic resonance imaging
Disciplines: Ingeniería
Keyword: Ingeniería eléctrica,
Bobinas,
Imagenología,
Resonancia magnética,
Simulación electromagnética,
Transceptores
Keyword: Engineering,
Electrical engineering,
Coils,
Imagenology,
Magnetic resonance,
Electromagnetic simulation,
Transceivers