Utilización de un dispositivo portátil para monitorear la relación internacional normalizada en pacientes con anticoagulante lúpico positivo



Título del documento: Utilización de un dispositivo portátil para monitorear la relación internacional normalizada en pacientes con anticoagulante lúpico positivo
Revista: Acta bioquímica clínica latinoamericana
Base de datos: PERIÓDICA
Número de sistema: 000435575
ISSN: 0325-2957
Autores: 1
1
1
1
1
1
1
Instituciones: 1Hospital Británico de Buenos Aires, Servicio de Hematología y Hemoterapia, Ciudad Autónoma de Buenos Aires. Argentina
Año:
Periodo: Dic
Volumen: 53
Número: 4
País: Argentina
Idioma: Español
Tipo de documento: Artículo
Enfoque: Analítico, prospectivo
Resumen en español La certeza del valor de la relación internacional normalizada (RIN), ensayo para controlar la anticoagulación con dicumarínicos, en pacientes con anticoagulante lúpico positivo (AL) es desconocida especialmente para los dispositivos al lado del paciente (POCT). El objetivo de este trabajo fue investigar si existe correlación entre el valor del RIN obtenido por el método tradicional y el obtenido con un dispositivo portátil en pacientes con AL positivo. Se estudiaron 35 pacientes anticoagulados por enfermedad tromboembólica con diagnóstico de AL positivo persistente a los que se les determinó al mismo tiempo el RIN por el método tradicional y con CoaguChek durante 4 controles consecutivos. El rango del RIN fue 1,9 a 5,60 y el RIN-POCT estuvo entre 2,0 y 4,92. La comparación del RIN vs RIN-POCT mostró r=0,98, pendiente: 1,56 (0,98-1,12) y una ordenada al origen de -0,088 (-0,282-0,007). El sesgo fue 2,1%. Para un nivel del RIN menor de 3,5 (n=136 controles) la diferencia del RIN promedio fue de 0,17 con un rango de 0,01-0,56. Un paciente, con triple positividad, mostró una diferencia entre ambos métodos mayor de 0,4 en dos controles. Para un RIN mayor de 4,5 el grado de concordancia fue menor pero no tiene implicancia clínica. Los resultados del RIN obtenidos por CoaguChek en los pacientes estudiados con AL positivo son útiles para la práctica clínica. Los datos obtenidos demuestran que hay una buena correlación entre el RIN tradicional y el CoaguChek. Por la gran diversidad de los equipos POCT los resultados no son extrapolables a otros dispositivos. Dada la heterogeneidad de los anticuerpos antifosfolípidos, es recomendable probar en cada paciente si hay una buena concordancia entre el RIN tradicional y el RIN-POCT
Resumen en inglés The certainty of the value of the international normalized relation (INR) assay to control dicoumarin anticoagulation in patients with positive lupus anticoagulant (LA) is unknown especially for the point of care testing (POCT). The aim of this work was to investigate if there was a correlation between the INR values obtained by the traditional method and those obtained with a POCT in patients with positive LA. The population under study were 35 patients anticoagulated by thromboembolic disease with a persistent positive LA, whose INR was determined at the same time by the traditional method and with CoaguChek during 4 consecutive controls. The INR range was 1.9 to 5.60 and the RIN-POCT was between 2.0-4.92. The comparison of INR vs. INR - POCT showed r=0.98, slope: 1.56 (0.98-1.12) and ordered to the origin -0.088 (-0.282-0.007). The bias was 2.1%. For an INR level lower than 3.5 (n=136 controls) the average INR difference was 0.17 with a range of 0.01-0.56. One patient, with triple positivity showed a difference between both methods greater than 0.4. in two controls. For INR greater than 4.5, the degree of concordance is lower but has no clinical implications. The data obtained show that there is a good correlation between the traditional INR and the CoaguChek. The results of INR obtained by CoaguChek in patients studied with positive LA are useful for clinical practice. Due to the large diversity of POCT, the results cannot be extrapolated to other devices. Given the heterogeneity of antiphospholipid antibodies, it is advisable to test in each patient whether there is a good agreement between the traditional INR and INR-POCT
Resumen en portugués A certeza do valor da razão internacional normalizada (RIN ou IIN), ensaio que controla a anticoagulação com dicumarínicos, em pacientes com anticoagulante lúpico positivo (AL) é desconhecida especialmente para os dispositivos de teste do tipo point-of-care (POCT). Este trabalho teve como objetivo pesquisar se existe correlação entre o valor de RIN obtido pelo método tradicional e aquele obtido com um dispositivo portátil em pacientes com AL positivo. Foram estudados 35 pacientes anticoagulados por doença tromboembólica com diagnóstico de AL positivo persistente aos quais lhes determinaram, ao mesmo tempo, a RIN pelo método tradicional e com CoaguChek durante 4 controles consecutivos. O intervalo de RIN foi de 1,9 a 5,60 e o de RIN-POCT ficou entre 2,0 e 4,92. A comparação de RIN vs RIN-POCT mostrou r=0,98, pendente: 1,56 (0,98-1,12) e uma ordenada à origem de -0,088 (-0,282-0,007). O viés foi 2,1%. Para um nível de RIN menor a 3,5 (n=136 controles) a diferença de RIN em média foi de 0,17 com um intervalo de 0,01-0,56. Um paciente, com tríplice positividade, mostrou uma diferença entre ambos os métodos maior a 0,4 em dois controles. Para um RIN de mais de 4,5, o grau de concordância foi menor, mas não tem consequências clínicas. Nos pacientes estudados com AL positivo, os resultados da RIN obtidos por CoaguChek são úteis para a prática clínica. Os dados obtidos demonstram que existe uma boa correlação entre a RIN tradicional e o CoaguChek. Devido à grande diversidade dos equipamentos POCT, os resultados não são extrapoláveis a outros dispositivos. É recomendável, visto a heterogeneidade dos anticorpos antifosfolípídes, provar em cada paciente a existência de uma boa concordância entre a RIN tradicional e a RIN-POCT
Disciplinas: Química,
Medicina
Palabras clave: Bioquímica,
Diagnóstico,
Química analítica,
Equipo médico,
Anticoagulante lúpico
Keyword: Biochemistry,
Diagnosis,
Analytical chemistry,
Medical equipment,
Lupus anticoagulant