EQA-based evaluation of metrological traceability of clinical chemistry test results in Argentina



Título del documento: EQA-based evaluation of metrological traceability of clinical chemistry test results in Argentina
Revista: Acta bioquímica clínica latinoamericana
Base de datos: PERIÓDICA
Número de sistema: 000435570
ISSN: 0325-2957
Autores: 1
2
1
Instituciones: 1Leiden University Medical Center, Leiden, Zuid Holland. Países Bajos
2Queen Beatrix Hospital, Winterswijk, Gelderland. Países Bajos
Año:
Periodo: Dic
Volumen: 53
Número: 4
País: Argentina
Idioma: Inglés
Tipo de documento: Artículo
Enfoque: Analítico, descriptivo
Resumen en español La equivalencia de resultados entre laboratorios es una mision importante para los laboratorios medicos. En los Paises Bajos, los laboratorios medicos solo usan aplicaciones comerciales homogeneas, regulatoriamente aprobadas (CE-IVD) para analitos quimicos, mientras que en la Argentina son comunes las aplicaciones heterogeneas caseras. El efecto de esta diferencia practica en la precision de la prueba se estudia utilizando caracteristicas clave del programa EQA, basado en la precision, de los Paises Bajos. Se ensayaron seis pools de sueros, congelados, de origen humano, conmutables, que cubrian el rango de medidas (pato)fisiologicas para 17 analitos de quimica clinica. Estos analitos de quimica clinica fueron analizados por ~75 laboratorios argentinos y ~200 laboratorios holandeses en 2014. Despues de eliminar los valores atipicos, el estado de armonizacion entre los laboratorios fue evaluado calculando los coeficientes de variacion interlaboratorios medios globales (CV%) por analito y por pais para los 6 niveles. No obstante, el estado de estandarizacion se evaluo despues de la eliminacion de valores atipicos mediante el calculo de recuperaciones medias generales (%) en comparacion con los valores asignados por analito por pais para los 6 niveles. Los sesgos medios absolutos se compararon con los sesgos (minimos / deseables) derivados de los criterios de variacion biologica. Para enzimas sericas los CV interlaboratorio en los laboratorios argentinos oscilaron entre 10 y 22%, en comparacion con 3-6% en los Paises Bajos. Para el acido urico serico, creatinina, glucosa y proteinas totales, los CV entre laboratorios variaron entre 4,3 y 13,1% en los laboratorios argentinos, en comparacion con <3,5% en los Paises Bajos. Para los electrolitos sericos, los CV interlaboratorios oscilaron entre 1,8 y 3,8% para Na+; 2,9-5,8% para Cl-; 3,8-7,5% para K+; 9,4-10,4% para Ca2+ y 16,2-22,3% para Mg2+ en comparacion a ≤2% (Na+, K+, Cl-, Ca2+) y ≤3% (Mg2+) en los Pais
Resumen en inglés Equivalence of results among laboratories is a major mission for medical laboratories. In the Netherlands, medical laboratories only use homogenous, commercial for general chemistry analytes, whereas in Argentina heterogenous, home brew test applications are common. The effect of this practice difference on test accuracy is studied using key features of the accuracy-based EQA program of the Netherlands. Six frozen, human-based, commutable poolsera, covering the (patho) physiological measuring range for 17 general chemistry analytes, were assayed by ~75 Argentinian labs and ~200 Dutch laboratories in 2014. After removal of outliers, harmonization status among laboratories was evaluated by calculating overall mean interlaboratory coefficients of variation (CVs, %) per analyte and per country for all 6 levels. Evenso, standardization status was evaluated after removal of outliers by calculating overall mean recoveries (%) as compared to the assigned target values per analyte per country for all 6 levels. Absolute median biases were compared to (minimal/desirable) biases derived from biological variation criteria. For serum enzymes interlaboratory CVs in the Argentinian laboratories ranged between 10 and 22%, as compared to 3-6% in the Netherlands. For serum uric acid, creatinine, glucose and total protein, interlaboratory CVs varied between 4.3 and 13.1% in Argentinian labs, as compared to <3.5% in the Netherlands. For serum electrolytes, interlaboratory CVs ranged between 1.8 and 3.8% for Na+; 2.9-5.8% for Cl-; 3.8-7.5% for K+; 9.4-10.4% for Ca2+ and 16.2-22.3% for Mg2+ as compared to ≤2% (Na+, K+, Cl-, Ca2+) and ≤3% (Mg2+) in the Netherlands. Mean recoveries in Argentinian laboratories for e.g. serum creatinine, glucose, CK, Ca2+ and Na+ were 95-119%; 95-104%; 98-102%; 98-102% and 96-100% respectively, whereas min-max recovery ranges were 65-155%; 58-126%; 47-132%; 66-132% and 85-115%. In the Netherlands, absolute mean recoveries were overall 98.9% wit
Resumen en portugués A equivalencia de resultados entre laboratorios e uma missao importante para os laboratorios medicos. Nos Paises Baixos, os laboratorios medicos so utilizam aplicacoes comerciais homogeneas, aprovadas por regulacoes (CE-IVD) para analitos quimicos, ao passo que na Argentina sao comuns as aplicacoes heterogeneas caseiras. O efeito desta diferenca pratica na exatidao do teste e estudado utilizando caracteristicas essenciais do programa EQA, dos Paises Baixos, baseado na exatidao. Foram ensaiados seis pools de soros, congelados, de origem humana, comutaveis, que abrangiam a faixa de medidas (pato)fisiologicas para 17 analitos quimicos gerais. Esses analitos quimicos foram analisados por ~75 laboratorios argentinos e ~200 laboratorios holandeses em 2014. Apos eliminar os valores atipicos, o estado de harmonizacao entre os laboratorios foi avaliado atraves do calculo dos coeficientes de variacao interlaboratorio meios globais (CV%) por analito e por pais para os 6 niveis. Nao obstante, o estado de padronizacao foi avaliado depois da eliminacao de valores atipicos pelo calculo de recuperacoes medias gerais (%) se comparados com os valores atribuidos por analito por pais para os 6 niveis. Os vieses medios absolutos foram comparados com os vieses (minimos / desejaveis) decorrentes dos criterios de variacao biologica. Para enzimas sericas, os CV interlaboratorio nos laboratorios argentinos oscilaram entre 10 e 22%, em comparacao com 3-6% nos Paises Baixos. Para o acido urico serico, creatinina, glicose e proteinas totais, os CV entre laboratorios variaram entre 4,3 e 13,1% nos laboratorios argentinos, em comparacao com <3,5% nos Paises Baixos para os eletrolitos sericos, os CV interlaboratorios oscilaram entre 1,8 e 3,8% para Na+; 2,9-5,8% para Cl-; 3,8-7,5% para K+; 9,4-10,4% para Ca2+ e 16,2-22,3% para Mg2+ em comparacao com ≤2% (Na+, K+, Cl-, Ca2+) e ≤3% (Mg2+) nos Paises Baixos. As recuperacoes medias em laboratorios argentinos para, p.ex. a creatinina ser
Disciplinas: Química
Palabras clave: Química analítica,
Bioquímica,
Química clínica,
Trazabilidad,
Argentina
Keyword: Biochemistry,
Analytical chemistry,
Clinical chemistry,
Traceability,
Argentina
Texto completo: Texto completo (Ver HTML) Texto completo (Ver PDF)