Viability and resolution of sensing hemolysis with a capacitive sensor



Título del documento: Viability and resolution of sensing hemolysis with a capacitive sensor
Revista: Revista médica del Hospital General de México
Base de datos: PERIÓDICA
Número de sistema: 000439514
ISSN: 0185-1063
Autors: 1
2
1
Institucions: 1Universidad Nacional Autónoma de México, Instituto de Ciencias Aplicadas y Tecnología, Ciudad de México. México
2Instituto Politécnico Nacional, Escuela Nacional de Medicina y Homeopatía, Ciudad de México. México
Any:
Període: Ene-Mar
Volum: 83
Número: 1
Paginació: 5-12
País: México
Idioma: Inglés
Tipo de documento: Artículo
Enfoque: Analítico, descriptivo
Resumen en inglés Monitoring lysis in cell’s suspensions can be a good indicator of cell’s integrity and health. The current techniques to measure the number of lysed cells in some process are generally time consuming. Objective: In this paper, we investigate the viability of a simple method for sensing hemolysis in real time through the electrical impedance of a parallel-plate capacitor that holds a biological cell suspension. Materials and methods: The method is based on measuring variations of the effective dielectric function of an erythrocyte suspension during a lysis process induced by osmotic stress. Our sensor is strongly affected by the so-called electrode polarization phenomenon which prevents sensing variations of the real part of the effective dielectric function. However, measurements of the imaginary part, and thus of the electrical resistivity of the suspension, is straight forward. Results: We show that from resistivity measurements, it is possible to sense variations of the lysed erythrocytes’ volume fraction with a resolution better than 0.1%. Conclusions: The proposed sensor can offer a high resolution in monitoring hemolysis and is a viable option for developing new rapid diagnostic medical tests
Disciplines Medicina
Paraules clau: Hematología,
Diagnóstico,
Suspensiones celulares,
Hemólisis,
Técnicas de diagnóstico,
Impedancia eléctrica,
Sensores capacitivos
Keyword: Hematology,
Diagnosis,
Cell suspensions,
Hemolysis,
Diagnostic techniques,
Electrical impedance,
Capacitive sensors
Text complet: https://www.hospitalgeneral.mx/frame_esp.php?id=47