Low noise low power readout for soft x ray detection



Título del documento: Low noise low power readout for soft x ray detection
Revista: Journal of applied research and technology
Base de datos: PERIÓDICA
Número de sistema: 000224010
ISSN: 1665-6423
Autors: 1

2
3
Institucions: 1International Center of Physical Theoretician of Trieste, Trieste, Friuli Venezia Giuli. Italia
2Instituto Politécnico Nacional, Centro de Investigación y de Estudios Avanzados, México, Distrito Federal. México
3University of Bucharest, Electronic Department, Bucarest. Rumania
Any:
Període: Oct
Volum: 1
Número: 3
Paginació: 187-193
País: México
Idioma: Inglés
Tipo de documento: Artículo
Enfoque: Aplicado
Resumen en inglés A new low power CMOS ASIC for the detection of X-rays was optimized for low power and low noise. Theoretical calculations and optimizations are presented and compared with experimental results. Noise as low as 120+25-Cin [pF] ENC rms was obtained including a silicon detector of 1.3 pF and 0.3nA of leakage. The power consumption is less than 100 W. Other circuit parameters are also shown
Disciplines Ingeniería
Paraules clau: Ingeniería de instrumentos,
Ingeniería electrónica,
Amplificadores,
Ruido,
Rayos X,
Detección
Keyword: Engineering,
Electronic engineering,
Instrumentation engineering,
Amplifiers,
Noise,
X-rays,
Detection